OPTIMIZATION, DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING RP-HPLC METHOD FOR SIMULTANEOUS QUANTIFICATION OF AMLODIPINE, OLMESARTAN AND CHLORTHALIDONE IN PHARMACEUTICAL FORMULATIONS
Jayprakash Sharma*, Dr. Sandeep Bhatia, Dr. Piush Sharma
ABSTRACT The present study aimed to develop and validate a simple, rapid, accurate, precise, and stability-indicating RP-HPLC method for the simultaneous estimation of Amlodipine, Olmesartan, and Chlorthalidone in combined tablet dosage forms. The optimized method used a C18 column (250 × 4.6 mm, 5 µm) with Acetonitrile and Potassium Dihydrogen Phosphate buffer (60:40, v/v) adjusted to pH 3.0 as the mobile phase. The flow rate was 1.0 mL/min, detection wavelength was 238 nm, injection volume was 20 µL, and the total run time was 10 min. The drugs showed well-separated peaks at retention times of 2.34, 4.58, and 6.92 min for Amlodipine, Olmesartan, and Chlorthalidone, respectively. The method was validated according to ICH Q2(R2) guidelines. Good linearity was obtained with R² values of 0.9997, 0.9998, and 0.9996, respectively. Recovery was between 99.42% and 100.36%, and precision values were within the acceptable limit of 2% RSD. Keywords: Amlodipine; Olmesartan; Chlorthalidone; RP-HPLC; Method Validation; Stability-Indicating. [Download Article] [Download Certifiate] |